Work with usRoom 7551 will hold 36 small hydroponic units. Each one runs as its own experiment, and an AI agent can set and read every one. This page shows how a unit works, what can fail, and what we need you to check.
The two existing racks stay where they are. Their long channels are cut into short ones and grouped four to a tray. We call the tray a cassette. A cassette slides out to the aisle onto a lift and carries no plumbing.
A pair is two cassettes back to back on one tier. They share a light frame and two fans. Each has its own water.
| Factor | Set per | Independent units | How |
|---|---|---|---|
| Nutrient strength and pH | Cassette | 36 | Dosing pumps hold a conductivity and a pH target |
| Ratio of stock A to stock B | Cassette | 36 | The two dosing pumps run in a set ratio |
| The stock solutions themselves | Bay of six cassettes | 6 | Each bay has its own jugs |
| Flow rate | Cassette | 36 | Pump voltage, 0.5 to 1.5 L/min per channel |
| Crop or cultivar | Cassette | 36 | |
| Day length and light level | Pair | 18 | A relay for on and off, dimming from 10 to 100 percent |
| Air speed over the leaves | Pair | 18 | Fan speed |
| Air temperature, humidity, COâ‚‚ | Room | 1 | Measured in every bay. Not controlled per unit |
Thirty-six loops give twelve nutrient treatments in triplicate. A light study has 18 units, so nine treatments in duplicate or six in triplicate.
Full power gives about 220 µmol/m²/s at the canopy, or 12.7 mol/m² over a 16 hour day. That is an estimate until the first pair is measured. Each pair can now be blacked out with motor blinds on its two aisle faces and fixed panels on its sides. The light trap for the fans is untested.
The 17 November meeting asked for channels in groups of four so that some plants are interior. One cassette does that with 16 plants. A study that needs more interior plants can run a pair as one plot.
We would like your view on the smallest plot you would accept for lettuce, cilantro and strawberry.
| Where | Measured | How often |
|---|---|---|
| Each tank | pH, conductivity, solution temperature | Every 10 seconds |
| Each tank | Low and high level | Continuous |
| Each cassette feed | Flow | Continuous |
| Each cassette side | Air temperature and humidity | Every 10 seconds |
| Each pair | Light on, fan running | Continuous |
| Each bay | COâ‚‚, leak | Every 10 seconds |
| Each rack | Water added to each tank, which is plant uptake | Per fill |
| Any cassette | Colour and near-infrared images | One shared camera, on request |
| Any tank | Handheld pH, conductivity, dissolved oxygen | Weekly check by a person |
The tank probes run the dosing. A handheld meter is the reference for all of them.
Each cassette also has a plugged port where the solution leaves the roots. A reference probe or a trial sensor can sit there. No such sensor is chosen yet.
Not measured: light spectrum, leaf temperature, single nutrients such as nitrate or potassium, and root images.
| Failure | What the design does | Still open |
|---|---|---|
| A pump stops | The flow sensor raises an alarm within 2 minutes. Only one cassette is affected. Four spare pumps are on the shelf. Each pump has a hand switch that works with the controller off. | Nobody is in the lab at night. An alarm needs a person who answers it. |
| One nozzle clogs | Each loop has a strainer ahead of the pump and a mesh sock on the return. | The flow sensor reads the whole cassette. One dry channel in four may not trip it. |
| A leak at a tier | A pan under each upper cassette drains back to its own tank. A tray under the tanks holds a leak sensor that shuts the fill valves. | |
| A fill valve sticks open | A float valve behind it closes. A high-level switch and a 10 minute limit close a second valve on the rack. | |
| A tank runs low | The low-level switch stops the pump and raises an alarm. The tank refills by itself in normal running. | |
| A power cut | Lights and fans go off. Batteries run every pump 1 minute in 5 for about 24 hours. | The safe gap between wettings. The run time is an estimate until tested. |
| The network or a computer stops | Each bay controller notices within 10 seconds, shuts its valves and keeps the water moving. Dosing pauses and resumes. | |
| A dosing error | Limits sit in the controller, outside the software: 20 seconds a dose, six doses an hour, never acid and base within 10 minutes. | |
| A probe drifts | A weekly check against the handheld meter. Recalibrate at 0.2 pH. | That is 72 probe checks a week. It is the largest routine job in the lab. |
| Root disease spreads | No loop shares solution with another. The feed crosses an air gap. A dye test on the first pair checks for carry-over. | Splash and airborne spread between neighbours are reduced, not ruled out. |
| Algae | Tanks are wrapped and covered. Channels have lids. | |
| The room overheats or stays damp | The racks fill in stages while room temperature and humidity are logged. | The lab adds about 7 kW of heat and up to 200 pints of water a day. Facilities has not said what the room can remove. |
| A rack rolls or tips | Levelling feet replace the casters. A pull test at 7 ft comes before first use. | A structural check. See section 9. |
| Cleaning falls behind | A cassette goes to the sink whole. A tank pumps itself out. | Thirty-six tanks and 144 channels to clean each crop cycle. |
The pumps, valves and controllers run at 24 V. Three power supplies feed them, and each charges its own battery. The lights run from wall outlets and are not backed up.
The load assumes each pump draws 25 W and runs 1 minute in 5. Both figures will be measured on the first pair.
The parts list sums to about $206,000 against $200,000. Every figure is an estimate. Nothing is quoted. The plan is to build and test one pair before ordering the other seventeen.